Objectives
To evaluate the long‐term course of quadripolar DBS electrodes with disconnected single contacts that cannot be used for DBS.
Materials and Methods
Quadripolar electrodes with open circuits of single contacts or monopolar impedances >6500 Ω were identified from a cohort of 2082 electrodes from 1044 patients with variable movement disorders. The long‐term course was analyzed from follow‐up data.
Results
Disconnected contacts were found in 58 electrodes (2.8%) from 49 patients (4.7%). The dysfunction was restricted to one contact in 51 electrodes (87.9%), two contacts in 5 electrodes (8.6%), three contacts in 2 electrodes (3.4%). Onset was related to surgery (implantation, impulse generator replacement, or other surgical revision) in 34 electrodes (58.6%), trauma in 2 electrodes, undetermined in 11 electrodes, and occurred spontaneously after previous normal measurements in 11 electrodes (19.0%). Repeated measurements at follow‐ups of ≥3 months were available in 39 electrodes. In 16 electrodes (41.0%) abnormal impedances persisted constantly during observations up to 11½ years (47 ± 35 months, median 41 months). In 21 electrodes (53.8%) abnormal impedances remained restricted to the initial contact(s) but varied considerably between measurements during up to six years (39 ± 18 months, median 38 months). Only two electrodes (5.1%) with initially one disconnected contact developed a disconnection of a second contact.
Conclusions
Disconnections of single contacts occur with increasing cumulative incidence during long‐term DBS. Surgery is the main causative risk factor. In the majority of electrodes, the dysfunction remains restricted to the initial contact(s).